Shoot-and-Dilute GC – A Way to Keep GC Systems Up Longer to Help Lower Cost of Analyses

Dilute-and-shoot LC-MS/MS is a selective and sensitive technique often used in determining pesticides in foods.  Because of competitive ionization effects seen with electrospray (ESI) and atmospheric pressure chemical ionization (APCI) methods, quantitative accuracy suffers for trace compounds of interest when matrix coelutions occur.  Dilution of extracts (or sometimes raw samples) is a good way of trying to mitigate this issue, although sample cleanup can help tremendously also.  For an excellent discussion of dilute-and-shoot LC-MS/MS, check out Julie Kowalski’s article in our latest Advantage newsletter, “Mitigating Matrix Effects: Sample Prep and Calibration; Strategies for Multiresidue Pesticide LC-MS/MS Analysis of Foods”.

Shoot-and-Dilute GC (aka “split injection”) is an analogous technique to dilute-and-shoot LC-MS/MS, used more because of matrix effects that can occur at the INLET and COLUMN, versus at the detector, although with ion trap mass spectrometers and the recently introduced, ultra-sensitive APGC systems, matrix effects at the detector end are still possible.  Rather than write a(nother) book on problems associated with splitless injection of dirty samples, I’ll simply mention that compound degradation and drastic response changes (especially for low volatility and active components) can quickly occur in a GC inlet liner that gets dirty.  And I’ll refer you to the extra reading below.

An easy way to mitigate the aforementioned problems is to use split injection GC any time you can.  That is, if you can meet LOD and LOQ requirements by making split injections at ratios of 10:1, 50:1, or even higher, increased flow through the inlet is going to minimize compound degradation (e.g. Endrin and DDT) and poor response for involatile compounds (e.g. PAHs, PCBs, dioxins and furans).  In addition to fg-detection ECDs, we have sensitive GCxGC-TOFMS and GC-MS/MS equipment in our lab, prime instrumental candidates for Shoot-and-Dilute GC.

I was going to say “it’s easy” to demonstrate the positive effects of Shoot-and-Dilute GC, and maybe it is, but it takes a LOOOOONG time because split injection keeps the GC system up so long versus splitless injection.  You can see that below in two chromatograms where I injected a Used Motor Oil Composite Standard hundreds of times, interspersed with a PAH Standard to monitor any response fall-off, and the system just won’t go down!  (I’m still running as I write this with no liner change!)  There is a slight baseline rise at the end of the run occurring over time (not shown because chromatograms are baseline corrected), likely from motor oil “heavies” eluting over several runs, but the responses for PAHs, even late eluting ones, are remaining very steady.  Pretty amazing, I think.

Consider your LODs, your LOQs, the sensitivity of your GC detector, the dirtiness of your sample extracts, grab a Restek Premium Precision split liner with wool and try Shoot-and-Dilute GC in your own lab.  You’ll save liners and columns, and money and time.

P.S.  The 15m x 0.25mm x 0.25µm Rxi-5ms GC column I’m using for this work ain’t no slouch either.  It’s low bleed out of the box and continues to have great lifetime while being hammered with used motor oil and held at 330°C over and over during this ongoing experiment.

Splitless injection GC extra reading:

Change the GC inlet liner – When?

Splitless injections of dirty samples on a single taper with wool liner result in quick degradation of its performance, especially for involatile polycyclic aromatic hydrocarbons

Setting the splitless injection purge valve time in gas chromatography

GC inlet liner choice made easy

The Impact of a Dirty Marijuana Extract on Pesticide Recoveries from a Splitless GC Inlet

Shoot-and-Dilute GC (split injection) extra reading:

Keep your GC organochlorine pesticide analysis systems up longer by using split injection.

Semivolatiles (including Appendix IX) on the Rxi-5ms by U.S. EPA Method 8270D (Split Injection)

Split Injection Minimizes Endrin and DDT Breakdown for GC-ECD Analysis of Organochlorine Pesticides

 

DS PAHs 02

 

 

 

4 Responses to “Shoot-and-Dilute GC – A Way to Keep GC Systems Up Longer to Help Lower Cost of Analyses”

  1. […] reactivity (e.g., captan breakdown) issues faced by residue chemists.  We call split injection, “shoot-and-dilute GC”, and are focusing our presentation efforts at RAFA 2015 on this technique, including two where we […]

  2. […] a recent ChromaBLOGraphy post I mentioned that we are doing a campaign on “shoot-and-dilute GC”, also known as split injection GC, at RAFA 2015. A paper just published by Tienstra, Portolés, […]

  3. […] P.S. If you’ve been reading my last few ChromaBLOGraphy posts, you’ve probably already guessed that the nitrogen 0.15mm column approach benefits by using “shoot-and-dilute” GC. […]

  4. […] a previous post, I introduced the Shoot-and-Dilute GC concept to keep GC systems up longer.   A split injection with a Restek Premium Precision® split […]

Leave a Reply


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us