Phthalate Determination by Dual Column Set in Eight Minutes

Dan Li, Rebecca Stevens, Jason Thomas and Chris English

Picture3

If you don’t have a GC-MS for identification of the EPA regulated phthalates, don’t worry! Here we introduce a parallel dual column set for the analysis of regulated phthalates using µECD. As shown in my previous blog , the Rtx-440 column (cat. # 12923) is a perfect choice for fast phthalate separation and quantification. The Rxi-35sil (cat. # 13823) is a good choice for a confirmation column. The dual column set can separate the 16 target phthalates listed in method EPA 8061A, including the internal standard benzyl benzoate (cat. # 31847), in less than 8 minutes

As seen in the chromatogram below, 4 pairs of phthalate peaks (in red) switched elution orders on the Rxi-35sil column, which provides confirmative information for identification. (Note: the peaks were numbered according to the EPA 8061A method.) Concentrations of the phthalates have been adjusted to get adequate and relatively equivalent response to the µECD.

blog figure1Columns: Rtx-440 30 m, 0.25 mm ID, 0.25 μm (cat. # 12923) and Rxi-35sil 30 m, 0.25 mm ID, 0.25 μm (cat. # 13823) using Rxi guard column 5 m, 0.25 mm ID (cat.# 10029) with deactivated universal “Y” Press-Tight® connector (cat.# 20405-261); Sample: EPA Method 8061A Phthalate Esters Mixture (15 components) (cat.# 33227), hexyl 2-ethylhexyl phthalate, benzyl benzoate (internal standard) (cat. # 31847); Injection: Inj. Vol.: 2 μL split ratio :50:1, Inj. Temp.: 280 °C; Liner: Restek Premium 4.0 mm ID Cyclo Double Taper Inlet Liner (cat.# 23310) Oven: Oven Temp: 150 °C (hold 1 min) to 330 °C at 30 °C/min (hold 2 min); Carrier Gas: He; Detector: μ-ECD @ 330 °C; Notes: Instrument was operated in constant flow mode (3.4 mL/min). This chromatogram was obtained using an Agilent μ-ECD. To obtain comparable results, you will need to employ a μ-ECD in addition to dual columns connected to a 5-meter guard column using a “Y” Press-Tight® connector. Concentrations are as listed.

Table blg

Compared to the method recommended by EPA 8061A using a 5-type/1701-type column set, the run time using the Rtx-440 / Rxi-35Sil MS columns has been greatly reduced from 40 minutes to 8 minutes. Additionally, the resolution has been greatly improved, especially for bis(4-methyl-2-pentyl) phthalate / bis(2-methoxyethyl) phthalate (peaks 5 and 6) and diamyl phthalate / bis(2-ethoxyehtyl) phthalate (peaks 7 and 8). The Rtx-440 column showed superior selectivity over 5-type columns resulting in a faster runtime.

Overall, we recommend the Rtx-440 / Rxi-35sil dual column set as an ideal solution for fast phthalate separation and identification. Due to flow rate limitations on some mass spectrometers, the short run time on GC-µECD may not be able to be reproduced on the GC-MS.

 

EPA8061AFigure1

Table 1blg

Figure 1 and Table 1 are from EPA Method 8061A . Column 1 is a 30 m x 0.53 mm ID x 1.5 µm 5-type column. Column 2 is a 30 m x 0.53 mm x 1.0 µm 1701-type column. Temperature program is 150 ºC (0.5 min hold) to 220 C at 5 ºC/min, then to 275 ºC(13 min hold) at 3 ºC/min.

One Response to “Phthalate Determination by Dual Column Set in Eight Minutes”

  1. […] 12 and 13), and bis(2-n-butoxyethyl) phthalate / bis(2-ethylhexyl) phthalate (peak 14 and 15). The Rtx®-440 and Rxi®-35sil columns are ideal as a parallel dual column set for electron capture d…Rtx®-440 and Rtx®-XLB columns showed the highest resolution in this fast analysis. Peaks that […]

Leave a Reply


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us