Fast Analysis of PAHs using Split Injection and a Short GC Column

Click on the this link:  Fast Analysis of PAHs  to see how I analyzed the EPA 16 polycyclic aromatic hydrocarbons (PAHs) in under 10 min using GC-MS by employing split injection with a Restek Premium Precision split liner with wool and a 15m x 0.25mm x 0.25µm Rxi-5Sil MS GC column.  Split injection allows a higher GC oven temperature start with good focusing for naphthalene, the most volatile PAH.  And the Rxi-5Sil MS stationary phase (5% phenyl type) is silphenylene, a backbone modification that provides a faster separation for benzo[b]fluoranthene and benzo[k]fluoranthene versus a typical 5% diphenyl column like the Rxi-5ms.

In addition to the speedy separation of the benzofluoranthenes, note the almost baseline separation for indeno[1,2,3-cd]pyrene and dibenz[a,h]anthracene.  Not bad!

Although you’ll see literature that encourages faster GC analyses through employment of smaller inside diameter columns, I prefer the 15m x 0.25mm x 0.25µm due to the better sample loading capacity for the 0.25mm x 0.25µm over something like 0.18mm x 0.18µm or 0.15mm x 0.15µm columns.  In addition, the larger ID column has better ruggedness than the smaller bore columns.

One more note on the use of 15m x 0.25mm x 0.25µm versus 30m x 0.25mm x 0.25µm columns.  Under vacuum-outlet conditions, i.e., using MS, the analysis time is about 3 times shorter for the 15m column.  You can prove that to yourself with the EZGC Method Translator and Flow Calculator.

Leave a Reply


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us