Looking for a Cryogenic Valve for your Shimadzu GC-MS?

I was working on a project for volatile sulfur analysis. It took so long to cool the GC oven to 30 °C, and yet the most volatile analytes were still difficult to separate. At that time, I realized that I needed a cryogenic cooling system for my Shimadzu GC-MS QP2010 plus!

However, I was not as excited as expected when I received this CRG-2010 N2 part. The CRG is designed to be installed to the left side of GC where my MS is located (see the picture below). Trade the CRG for MS? That would not work for me.

manual pic

With the help of my colleagues, Becca and Corby, we decided to attach the CRG to the top of the GC, which turned out to be even better, because it is closer to the electronic board and it has a shorter route to the nitrogen tank too (see picture 1).

The installation can be as simple as these 5 steps:

pic 1 4

  1. Introduce the nozzle into the oven by piercing through the top insulation and then connect the copper extension tubing into the oven (picture 2 and 3).
  2. Mount the side plate of the valve to the top board of the instrument (you may need to drill a hole for screw).
  3. Plug the cable into the side electronic board (picture 4).
  4. Connect the nitrogen tank to the valve.
  5. Assemble the outer box.

Picture2Picture3

 

 

 

 

 

 

 

 

 

If you don’t have other detectors (like the FPD that I have installed), the back position is another good option. In this case, you don’t need to drill hole for screw (Picture 2).

Leave a Reply


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us