Cannabis Residual Solvents Using MS Detection – I’m Not Hungry but I’ll Eat My Words Anyway

After coming back from a huge lunch at the Bellefonte Wok, a favorite Restek lunch spot, I’m completely stuffed, but I have to eat my words from a previous blog. In this blog, I made the case against using MS detection for headspace analysis of residual solvents in cannabis concentrates due to interference between the air peak from the headspace injection and propane. It turns out that’s absolutely not the case, as my colleague and Restek air chemist Jason Herrington argued when I first posted the blog.

During a trip to Trace Analytics in Spokane, WA, I had the opportunity to run cannabis residual solvents using the full evaporation technique with headspace GC/MS (FET-HS-GC/MS), and lo and behold, propane is resolved from the air peak as shown in Figure 1.

Figure 1: Propane is Resolved from Air Peak on the Rxi-624Sil MS using FET-HS-GC/MS (25ppm Standard)

Figure 1

Propane is still well-resolved even at higher concentrations where band broadening may become an issue (Figure 2):

Figure 2: Propane is Resolved from Air Peak Even at Higher Concentrations (500ppm Standard)

Figure 2

Given the close proximity of propane to air, the two peaks may co-elute under sub-optimal injection conditions.

Although I dismissed the use of MS originally, there are benefits that I failed to mention. In addition to the ability to confirm peak identity using mass spectra, sensitivity is improved over GC-FID for the later eluting compounds like benzene, toluene, and the xylenes (BTX). This is because these compounds produce higher molecular weight fragments (78, 91, 106 m/z), and MS detectors aren’t great at reliably detecting low mass fragments from solvents like propane, methanol, and butane (29, 31, 43 m/z). Luckily, the solvents that produce the lower molecular-weight fragments have much higher regulatory cutoffs than butane, toluene, and xylenes, so detectability via MS isn’t an issue. Figure 3 shows an extracted ion chromatogram of our low standard (0.5ppm) and the signal-to-noise ratio is very good for BTX. Note that this chromatogram was collected in scan mode, so even more sensitivity can be gained from the development of a selected ion monitoring (SIM) method.

Figure 3: Good Sensitivity is Achieved for BTX using FET-HS-GC/MS (XIC of 78, 91, 106m/z, 0.5ppm Level)

Figure 3

All this being said, there is one big caveat to the use of MS for analysis of residual solvents in cannabis concentrates: the limited linear dynamic range of MS detectors when compared to FIDs. Where a FID can produce a linear curve over several orders of magnitude, an MS detector has a dynamic range limited to about three orders of magnitude. This means that one curve covering the entire regulatory range (e.g. 0.5 – 5500ppm) cannot be run using MS. In fact, under the conditions listed in the protocol for this method, BTX ions begin to saturate the detector at 250ppm (Figure 4), and almost all major ions for all analytes are saturated at 500ppm (Figure 5). This results in quadratic calibration curves for our analytes over even this limited range of quantification (Figure 4 inlay). If we tried to extend the curve beyond 500ppm, even a quadratic fit would be inappropriate.

Figure 4: BTX are Saturated at 250ppm, Resulting in a Quadratic Calibration Curve (Inlay)

Figure 4

Figure 5: Almost All Compounds are Saturated at 500ppm

Figure 5

While quadratic curves aren’t inherently bad, most guidance suggests the use of a linear calibration curve if feasible. Since even a quadratic curve wouldn’t cover the full regulatory range, in the cannabis residual solvents analysis protocol I suggest breaking up calibration into high and low segments. These segments will vary by state due to different regulations, but the approach is the same. For some states with very high cutoffs (I’m talking to you, Oregon), a higher split will most likely have to be used for the higher range curves in order to avoid overloading the column. Please note that this protocol has been partially validated, but still needs some work in terms of internal standard choice and extending the range of quantification above 500ppm.

While I learned a lot in the development of this method, I think the most valuable lesson I learned is to never argue with the air chemist when it comes to volatiles analyses.

4 Responses to “Cannabis Residual Solvents Using MS Detection – I’m Not Hungry but I’ll Eat My Words Anyway”

  1. Jack Cochran says:

    The “full scan” calibration range of any MS system can easily be extended by picking a less intense m/z ion for the upper level calibration points, essentially establishing two calibration curves, while only using one GC method (rather than changing the split ratio) and making one run per calibration standard. An example would be employing the 77 or 92 m/z ions for xylene calibration at higher levels (rather than 91 or 106 m/z ions, which are often used for xylene calibration/quantification). When I worked at LECO, we built this functionality into ChromaTOF, the software platform, and called it “Extended Range Calibration”. After simple setup, the software did all the work for generating the calibration curves and picking which one to use for quantification of samples.

  2. Jason S. Herrington says:

    The air/propane argument is water under the bridge. However, I am not sure I can get over the fact that you were at the Wok without me.

  3. All very good points, and we thank Restek for their continued support of this fledgling industry. It would also be good to hear from someone regarding the presence of endogenous solvents being produced by the plant itself. Our laboratory routinely finds small organics like acetone, ethanol, methanol and isopropanol in both concentrates and plant materials, much to the chagrin of our clients. From the literature, we have found that trace levels (sub 50 ppm) are typical from plants, and we can expect that these may be concentrated during a cannabinoid extraction. Now it’s just a matter of getting regulators to recognize this…

  4. Ini says:

    Is it possible for you guys to send me that partial validation data on the “Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates?”

Leave a Reply