GC carrier gases, do you really have a choice?

With the ongoing helium shortage, there are still plenty of analysts looking for alternative GC carrier gases.  So what other options are out there, and are they really viable?  In this post, I will briefly discuss the two most common alternatives (hydrogen and nitrogen), and when they really can’t be used as an alternative to helium.

 

N2

Let’s start with nitrogen.  This inert gas was seldom considered because in order to obtain maximum capillary column efficiency, carrier gas linear velocities (flow rates) needed to be much slower than either helium or hydrogen, which led to longer sample analysis times.  Longer analysis times meant fewer samples getting analyzed which led to less revenue.  As a result, using nitrogen as a carrier gas was usually reserved for packed columns.

 

H2

In contrast, hydrogen was used more often than nitrogen when using capillary columns because carrier gas linear velocities (flow rates) could be much higher (approximately 4X nitrogen and 2X helium) and still maintain capillary column efficiency.  As a result, shorter analysis times were possible.  However, hydrogen has several drawbacks of its own, with the two most common being safety and reactivity (hydrogen is not an inert gas).

 

He

It’s not surprising that helium became the preferred GC carrier gas.  It is inert, safe, and provides high column efficiency at moderate linear velocities (flow rates).

 

So why did I title this post “GC carrier gases, do you really have a choice?”  It’s because, in some cases, you really don’t have a choice when it comes to choosing a carrier gas.  For example:

 1.  Many EPA methods list helium as the carrier gas, so if you want to follow the method as written, you don’t have a choice.  The same is true for many other methods, including those from OSHA and NIOSH.

2.  Depending upon the GC detector, you may not have a choice.  This is especially true with “universal” detectors.  Consider a helium ionization detector (HID), a pulsed discharge detector (PDD), a pulsed discharge helium ionization detector (PDHID), and a discharge ionization detector (DID).  These all need helium to operate properly.  Before you try using a carrier gas other than helium, you should contact the detector manufacturer.

3.  Continuing on the topic of universal detectors, you may not have a choice of carrier gas if using a TCD (thermal conductivity detector).  Consider the following:

a.  You cannot choose a carrier gas which is a target compound.  For example, you cannot use nitrogen as the carrier gas if nitrogen is a target compound.

b.  You should choose a carrier gas which is the same as the balance gas (in your calibration gas standard and/or samples).  For example, if your balance gas is helium, you should use helium as your carrier gas.

c.  The carrier gas you choose for an analysis will affect target compound responses.  Valco (VICI) has an excellent Carrier Gas Selection section in their TCD manual (pages 7 & 8).   I pasted their table on thermal conductivity values below.  The important thing to remember is that, generally speaking, the larger the difference in thermal conductivity values between the carrier gas and target compounds, the more sensitivity for the target compounds.

Gas / Thermal Conductivity

 Hydrogen         45.9

 Helium             36.9

 Neon               11.8

Methane            8.6

Oxygen             6.6

Air                    6.4

Nitrogen            6.4

CO                   6.2

Water                4.5

Argon               4.5

CO2                  4.2

4.  When analyzing for hydrogen using a TCD, you should probably not use helium as the carrier gas. You can read more about this topic in the following links.

Thermal Conductivity Detector Analysis of Hydrogen Using Helium Carrier Gas and HayeSep® D Columns

Hydrogen Detection with a TCD using Mixed Carrier Gas on the Agilent Micro GC

 5.  Finally, if you are using the most popular GC universal detector, mass spec, a variety of reported issues may prevent you from using any carrier gas other than helium.  I covered several of these issues in a post I wrote last year.  It’s a confusing time for GC/MS analysts

 

So if none of the examples listed above are relevant to your analysis, then you probably do have a choice when it comes to carrier gas.  However, if your analysis fits into one of the examples listed above, you may not have a choice.

To read more about GC carrier gases, please see other posts written in the Restek ChromoBLOGraphy category titled Alternative GC Carrier Gases.   Thanks for reading.

2 Responses to “GC carrier gases, do you really have a choice?”

  1. Dear Alan –

    The response of gases (CO, CO2, CH4, N2 and O2) by TCD are actually quite good, when using argon as carrier gas. The TCD response depends not only on differences in thermal conductivity between the carrier gas and the analytes. Thermal capacity of the gases and the molar flow through the detector do also have major influence on the TCD response.

    With kind regards –
    Lars Kürstein, Copenhagen

  2. Alan Sensue says:

    Hi Lars:

    Thanks for reading my post. I am by no means an expert on TCD’s and compound responses when using argon as the carrier gas, so I truly appreciate your valuable feedback. I’m sure our customers will benefit from it.

    Alan

Leave a Reply


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us