Discovering Analyte Breakdown On-Column

I completed my winter internship at Restek and my main task was to expand the Pro-EZGC libraries.
So I created a library for Nitrosamines on the Rxi-1301Sil MS and while acquiring data we noticed a foot on a couple of my peaks.

Figure 1: On-column breakdown can be observed for the late eluting nitrosamine compounds as indicated by the foot at the leading edge of the peak.

I attempted to translate the simulation to a column with the same phase, but with a thicker film. Since the Rxi-624Sil MS and Rxi-1301Sil MS are the same stationary phase, I modeled a run with the Rxi-1301Sil MS Library using the column dimensions of an Rxi-624 Sil MS. As I ran the conditions provided by EZGC we noticed a strange rise of the baseline.

Figure 2: Rise in the baseline is observed which is most likely caused by analyte breakdown on-column. Breakdown components remain in the stationary phase and are observed as carryover.

One of my colleagues immediately suggested that the nitrosamines could break down as a function of elution temperature. In order to test this theory, I prepared three different run conditions with different elution temperatures for the late eluting nitrosamines on the Rxi-624 Sil MS:

Figure 4: Adjusting the final hold changes the elution temperature of the last three compounds. If the compound response drops as elution temperature increases it will support our theory of on-column breakdown.

These are the compounds I used (sorted by elution order):

Figure 5: List of nitrosamines found in the chromatogram where the last three compounds are labeled as 1, 2 and 3.

Figure 5: Three chromatograms illustrating the three different programs with increasing elution temperature of the last three nitrosamines. Notice the significant increase in breakdown.

I labeled the last three peaks; the ones that break down appear to have the most activity on-column. Thanks to the MS I was still able to determine the retention times and it became evident that the breakdown increases with higher elution temperature. Was this a general problem with these compounds on this phase, or was this specific to the Restek column?

 

Figure 6: Chromatograms with three different oven conditions. The last three nitrosamines have increasing elution temperatures from top to bottom. These chromatograms were performed on a competitor column to verify the on-column breakdown is most likely related to the cyano stationary phase and not specific to a manufacturer.

We only had a 60m competitor column available, so it’s not a direct comparison, however, we see similar breakdown. We can tell that the nitrosamines break down on-column on this phase in general. But if you use a film that is thin enough and run the compounds with a low elution temperature it will minimize on-column breakdown. Using an Rxi-1301Sil MS is still faster than using a “5-type” phase, but you will have to enter your nitrosamines into ProEZGC and find out for yourself.

Leave a Reply


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us