FDA Warns Hand Sanitizers May Contain 1-Propanol

The Food and Drug Administration (FDA) has expanded their recall of hand sanitizers and found more chemical contaminants to include 1-propanol (n-propanol cas# 71-23-8). 2-propanol (isopropanol cas# 67-63-0) and Ethanol (cas# 64-17-5) are the two most widely used active ingredients in hand sanitizers. With this latest addition to the list, the main concern for exposure is ingestion (1). In a three-year period over 60,000 children under 12 years old ingested hand sanitizer (2). 1-propanol is metabolized via the enzyme alcohol dehydrogenase to propionic acid and thereby disrupts the pH balance of the blood, while 2-propanol is metabolized with the same enzyme, however, the product is acetone leading to ketosis. Ingestion of both alcohols results in 1-propanol slowing down the breakdown of 2-proponal to acetone prolonging the sickness (3).

Our previous blog addressed the recall of hand sanitizers that may contain methanol. Using the same method conditions it’s possible to analyze 1-propanol in addition to other compounds commonly found in adulterated hand sanitizer (Figure 1). Following the conditions below, we recommend eliminating 1-propanol as a co-solvent in the rinse vial and substituting acetonitrile instead. Figure 2 shows 1-propanol can be found as a contaminant in the blanks and samples.

Figure 1: The analysis of denatured alcohol with an overlayed chromatogram of 1-propanol. It is possible to analyze for a variety of alcohols in hand sanitizer at percent levels using the Rtx-VMS.

Figure 2: Trace amounts of 1-propanol can be found in samples and blanks when using 10% 1-propanol as a post rinse (see red text). Changing the co-solvent to a non-target compound is recommended.

https://www.fda.gov/drugs/drug-safety-and-availability/fda-updates-hand-sanitizers-consumers-should-not-use 
2 https://www.sciencedirect.com/science/article/pii/S0099176718304884

3 https://pubmed.ncbi.nlm.nih.gov/18375643/

Related Reading Links:
https://blog.restek.com/swafs-chromatography-of-beer-urine/

Leave a Reply


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us