Molecular Sieve Packed Columns and Fixed (Permanent) Gas Analysis

For some reason, I find molecular sieve packed columns interesting.  Maybe it’s because they have a niche as the go-to columns for fixed (permanent) gas analysis.  Maybe it’s because they are old school technology that a new generation of analysts are always asking questions about.  Maybe it’s because using them successfully is a function of skill, art, and some luck.

As I researched molecular sieve packed columns over the years, I realized that for something that has been around more than 50 years, there wasn’t a whole lot of published information about them (at least not that I could find).  Application chromatograms were few and far between, and troubleshooting information seemed even rarer.  As a result, I decided to write this post on molecular sieve packed columns to help our customers.  This information came from books, the internet, coworkers, or was provided by our customers.   If you have any additional or contradictory information, I would love to hear from you.

I have written two other posts on molecular sieve packed columns.  One was titled “ShinCarbon columns – Installation / Conditioning / Helpful Hints”, and the other “Molecular Sieve 5A & 13X packed columns – Installation / Conditioning / Helpful Hints”.   So what’s the difference between these columns?  Let’s start with Molecular Sieve 5A and 13X packed columns.

.gc_pc00166

Click on Chromatogram to Enlarge

.

Molecular Sieve 5A and 13X

Molecular Sieve 5A and 13X are referred to as zeolite molecular sieves.  These crystalline materials are synthetically produced and have very uniform internal cavities.  I am not going to discuss structural/composition details of Molecular Sieve type A materials verses Molecular Sieve type X materials, but rather focus on helping new analysts understand these packing and their applications a little better.

So what is the difference between 5A and 13X?  Well, as I already mentioned, material structure is one difference.  Pore size is another; a 5A has an effective pore size of 5 angstroms and a 13X has an effective pore size of 10 angstroms.  In order for a compound to be able to pass through these pores, it must be smaller.  For the compounds I plan to discuss (He, H2, Ar, O2, N2, CH4, and CO), each is able to pass through these pores.

So which compounds won’t pass through these pores?  Well, I’m going to answer a different question and list several common gaseous compounds which should not be introduced onto these columns.  These compounds include CO2, H2O, SO2, H2S, Cl2, HCl, NH3 (and other corrosive vapors), compounds larger than (and including) C2 (ethane) and/or compounds with boiling points greater than approximately 50C (please note: there are several high temperature hydrocarbon applications, but I will not discuss them here). Larger gaseous compounds may actually make it through the column, but many will have no retention because they cannot fit into the pores (like SF6 when using 5A).  Other larger compounds may simply not make it through the column.

So what are some of the advantages of using these columns for gas analysis?  The primary advantage is baseline separation between O2 and N2 with a relatively short column and at room temperature.  Another is the separation of CO from air and CH4 (a difficult separation for any non-molecular sieve packed column).

So what are the disadvantages of these columns?  The primary disadvantage is that these packings will trap CO2 and H2O, which is why these columns are typically used with a pre-column (stripper column) and valve switching.  However, if CO2 and H2O do get trapped in these columns, the effects are reversible (see “Molecular Sieve 5A & 13X packed columns – Installation / Conditioning / Helpful Hints”).

Although zeolite molecular sieve packings are commonly used for argon/oxygen separation, a disadvantage of using a packed column is that a very long length (approximately 24ft from what I have read) is needed to separate oxygen from argon at room temperature.  For (O2/Ar) separation, we usually recommend one of our Molecular Sieve 5A PLOT columns (click on chromatogram below to enlarge).

gc_pc1182

gc_pc1182

gc_pc1182Finally, one needs to be cautious when using any carrier gas other than nitrogen to dry/condition these columns.  Because these materials contain surface metals that cannot be easily removed, it is believed that other carrier gases like hydrogen, argon and helium (at high temperatures) can remove the oxidation layer on the metal’s surface, which can lead to activity (especially for CO).

 

ShinCarbon ST

ShinCarbon ST is a synthetically manufactured carbon molecular sieve material with a very high surface area of approximately 1500m2/gm and a pore size of 10-12 angstroms.  Its surface is very inert, and very non-polar.  It is typically used for separation of fixed gases and light hydrocarbons, but also can be used for the analysis of % levels of SO2 (usually elutes after ethane) and ppm / % levels N2O (usually elutes after CO2 but before acetylene).

So how does a carbon molecular sieve column (ShinCarbon) compare to a zeolite molecular sieve (5A & 13X) column?  Let’s start by looking where it doesn’t quite match-up with a 5A or 13X.  First, it can be difficult to get baseline separation between O2 and N2 using a carbon molecular sieve column (customers report that a 10ft x 1/8” OD column is needed).  In addition, it is even more difficult to separate O2/Ar (I have read that cryogenic cooling is a necessity).  And just like the 5A & 13X, a ShinCarbon isn’t meant for H2S or any corrosive gases like Cl2, HCl, NH3.

gc_pc00666

Click on Chromatogram to Enlarge

So why would you use a ShinCarbon?  First, unlike zeolite molecular sieves, CO2, SO2 and H2O will not get trapped in the pores (Notes: it will be difficult to quantitate H2O because the peak will tail; there are better column choices for SO2).  Plus, you can easily chromatograph larger compounds (like straight-chain hydrocarbons up through and including pentane).  However, some unsaturated hydrocarbons (in particular, C3+) may show a lower response than expected, especially at higher temperatures.  Hexanes will elute but tail.

If H2O does condense inside the column, 30 minutes at 250⁰C will dry a ShinCarbon (unlike the 5A & 13X which needs 300⁰C for 3 hours), and using a carrier gas other than nitrogen does not seem to add surface activity.  Personally I do not suggest leaving it at higher temperatures (>250⁰C) for more than 3-hours (and certainly not overnight) since some customers have reported that the packing could develop surface activity under these conditions.

 

To view several ShinCarbon chromatograms using 80486-___ (our 1/8″ OD column), click here Shincarbon 80486

To generate the first chromatogram, 200µl of our catalog # 34543 was injected.

To generate the second chromatogram, 50µL of air was injected.

To determine the complete catalog number to purchase a 1/8″ OD Shincarbon column (including the required suffix number), please review this link Things to Consider Before Ordering a Packed Column

 

I hope that you have found this information useful.  As I mentioned, if you have any additional information, or you had an experience which is contradictory to something here, please let me know.  Thanks for reading.

35 Responses to “Molecular Sieve Packed Columns and Fixed (Permanent) Gas Analysis”

  1. Dear Alan –

    I was introduced to the MS5A and MS13X packed columns back in 1989, when I started working with gas chromatography. At my company we now have 50 GC’s configured with packed columns for gas analysis, including lots of moleculare sieves. Easy to build. Easy to operate. Excellent for separation of permanent gases, also in combination with other packed columns, like the Porapak and HayeSep. Precolumns for backflush and valve options for MS column isolation are mandatory. We actually wash the MS13X material for the RT change of CO and CH4! Total syngas analysis in seven minutes at 40C isothermal are easy achieved, using the nice Porapak-N and MS13X in combination. You can use He/H2, Ar, N2, H2 and He as carrier gas, depending on the application and component response at the TCD.
    From time to time, we think about changing the packed columns to PLOT columns. But we stick to the packed columns, due to larger column capacity on packed columns and lower TCD detection limits.
    I am wondering why we not see any PLOT-N and PLOT-13X capillary column? Are they difficult to build?

    With kind regards –
    Lars Kurstein
    Copenhagen, Denmark

  2. Jaap says:

    Hi Lars:

    The PLOT N is basically a Rt U-BOND. Its a high polar porous polymer and it is available on fused silica nd also on MXT.

    The 13X is used for gases , but for this separation, the 5A can do everything that is required. The 13X only adds only to a shorter retention for CO, which can also, be achived by running the 5A at 15-30C higher temperature

    The 13X has its own unique application for N-P separations of C1-C12 hydrocarbon mixtures.

    I do not know what you mean with ‘Washing” 13X columns. If you hydrate molsieves, they get less polar and CO will come off fast. The column is however containing a lot of water which is a recipie for reention time changes.

    regards
    jaap

  3. Ivy Soon says:

    Hi, this blog is really good that I’m starting to read its contents and learning a lot. This is the closest to what I do and I have a question. I’m have Molsieve 13X with TCD (in tandem with Hayesep N with FID) for TOGA, using direct injection with valve timing and I get a ghost peak after O2, There’s quite a time difference for O2 and N2 RTs and this ghost peak almost always has the same area with H2. I don’t know how to fix it. I used a calibration gas to do test runs. Could it be standard contam? or I have to adjust the valvel timing? Any input would be much appreciated. Cheers!

    Ivy

  4. Alan Sensue says:

    Just to let our readers know, I’ve been working with Ivy for the past few days on understanding her instrument set-up and trying to determine what is going wrong. I believe valve timing is part of the issue, but column selection is another. Her 8ft HayeSep N will not separate O2, N2, argon and CO under normal analytical conditions, but she wants to send CO (and methane) to a FID to obtain lower detection limits (she has a methanizer which converts CO to CH4), and nitrogen & O2 to a TCD. Unless cryo cooling is incorporated, or a different pre-column is selected, I don’t see how she is going to get this to work. My best guess is that she may need to use a ShinCarbon instead of the HayeSep N.

  5. Scott McFarlane says:

    This looks like a good candidate for a simple two column injection with a 10-port. The 10-port can be plumbed with two sample loops to simultaneously inject onto two different column/detectors. MS5A/TCD for O2, N2, and Ar; porous polymer/FID for the CO and CH4. It’s probably not a bad idea to add a pre-column backflush setup on the MS5A channel, which would require an additional valve.

    If this is trace analysis, the ghost peak might be a result of leakage in a valve. I don’t know what the setup is here, but we’ve seen that sort of thing when there is a 4- or 6-port valve in between two columns. Unlikely, since it’s a TCD and such leaks are usually in the PPB range. Typically there should be two ghosts as well, one prior to N2 and one after. Standard contamination can almost certainly be ruled out. Nothing that I’m aware of should be eluting between O2 and N2 on a MS column.

    Hope that helps!

  6. Alan Sensue says:

    Hi Scott:

    Thanks for reviewing my post and the comment. Ivy had sent me a follow-up email showing me a much improved chromatogram. It appears the analysis is working much better. I don’t remember the details, but I do believe there was a small leak which was causing the ghost peak. I do remember that the balance gas was nitrogen, and the carrier gas was helium. I usually caution our customers that when using a TCD, it is usually better to use the same carrier gas as the sample/standard balance gas. If anything, it eliminates one very large, unnecessary peak from the chromatogram.

    Thanks again for the input.

    Alan

  7. QB says:

    Hello Alan,

    Thank you for your post. It helps a lot for someone like me who is new to packed columns.

    I have a question in terms of Helium analysis by GC-TCD using a packed column. Zeolite molsieve vs. ShinCarbon, which column is a better choice? Currently, I can detect 1ppm Helium in Argon by using a Zeolite Molsieve 13x column. I am wondering if I can achieve lower detection limit with a shincarbon since it has large surface area and slight larger pore size?

    Thank you,
    QB

  8. Alan Sensue says:

    Hello QB:

    I am glad to hear that you have found my blog post helpful. Thank you for letting me know.

    I am surprised to see that you can detect 1ppm with a standard TCD. Usually detection limits for these detectors is triple-digit ppm, and not single-digit. I have seen however, that there are µTCDs which have much lower limits (some advertise single-digit ppm levels).

    Since compound detection limit is primarily a function of the detector for this type of analysis, as long as your peak shape looks good, I would not expect you to observe much difference in sensitivity by switching columns.

    Alan

  9. RAUL LAPINSKAS says:

    Hello QB,

    I am also surprised you can quantify 1 ppm of helium, using a TCD/Argon.

    I analise helium in several gases, by TCD/argon carrier/MS 5A 12’x 1/8″ column.

    Carrier flow is 30 mL/min, 60°C/ loop=1 mL.

    I got a LOQ=4 ppm(2 x noise).

    But, take care: Neon peak elutes very close to helium , and may be mis-identified.

    In my analysis above, RT for Helium is 0,82 min, and Neon elutes at 0,86 min.

    I think Shincarbon column is not recomended for this analysis, because resolution for helium,neon and hydrogen is far less than MS.

    Even MS 13X may be not suitable, because resolution is not as good as MS5A.

  10. Sue says:

    Hi Alan
    Thank you for the helpful hints on Molecular Sieve Packed Columns.

  11. Alice says:

    Hi Alan,

    Really useful blog – thank you. I am relatively new to gas analysis by GC (I previously worked with liquid samples and GC-FID), but am trying to improve our system a bit as we have problems resolving O2 and N2 peaks in emissions from soil samples – I would be very grateful if you have any thoughts. We have a fully enclosed system that enables us to replace the soil atmosphere with a He/O2 (80/20) mix and determine soil N2 production. We currently use a molecular sieve 5A PLOT column, He carrier gas and an HID detector. We do not have any pre-column scrubber installed. Whilst we can resolve O2 and N2, the separation is quite poor and declines over time (presumably due to moisture build-up in the column?) until the N2 peak is eventually absorbed into the O2 peak. Chromatographic performance can then be improved again by column conditioning. It does not help that O2 concentrations are high compared with N2 concentrations. The only other gas we occasionally wish to detect in this (part of the) system is methane. I am wondering whether there is anything I can do to improve things (e.g. perhaps by using a different column) and would be very grateful if you have any suggestions? Please let me know if any further details would be useful.
    Many Thanks,
    Alice

  12. Nat says:

    Hi Alan,

    In this blog you say ‘So which compounds won’t pass through these pores? Well, I’m going to answer a different question and list several common gaseous compounds which should not be introduced onto these columns. These compounds include CO2, H2O, SO2, H2S, Cl2, HCl, NH3’

    Could you please expand on why these gasses shouldn’t be introduced to the column, in particular I am interested in CO2.

    Many thanks,
    Nat

  13. Alan Sensue says:

    Hi Nat: I always assumed it was the shape of the molecule that caused it to get stuck (trapped) in the pores of the particle. Maybe one of our readers knows for sure. Alan

  14. Alan Sensue says:

    I had asked Alice for more information, and she sent it to her local Restek Office. It appears that high amounts of CO2 were to blame.

  15. Dave says:

    Hi Alan,

    I am quite new to the use of molecular sieves and want to follow up on Nat’s question (23 Mar 2017) and your response (24 Mar 2017). If CO2 gets trapped, will other gases still be able to travel through albeit slowly, or does the trapped CO2 effectively prevent any other gases from making it through?

    Thanks in advance (and thanks for a very informative blog).

    Dave

  16. Alan Sensue says:

    Hi Dave:

    Thank you for your question.

    My understanding is that the trapped CO2 prevents other compounds from entering the pores, which reduces the separation power (theoretical plates) of the column. However, I cannot say with absolute certainty that this is what happens. If you need a definitive answer, I suggest discussing this with a materials scientist; I would think they would know, or at least know where to look for the answer.

    Regards,

    Alan

  17. Dave says:

    Alan,

    Thank you!

    Dave

  18. Charles says:

    Hi Alan

    Can I use the Molsieve 13X I want to test for H2 in ppm Level my samples will be 99.5% Propylene and Ethylene.

    Regards,

    Charles

  19. Alan Sensue says:

    Hello: Because I normally advise against putting C2 hydrocarbons (ethane, ethylene, acetylene) onto a 5A or 13X column, I would suggest a carbon molecular sieve column for this particular analysis, like our ShinCarbon. Regards, Alan

  20. Sanni says:

    Hi Alan,

    You wrote this in 2013 and I was wondering if there have been any advancements made since your initial posting? Considering the importance to some industries, I suppose there are still people searching for a stationary phase that unlike 5A is unpolar and experiences no detrimental effects from water but can also separate with shorter retention times/column lengths than the ShinCarbon ST column apparently does?

    Most publications I found work with either a filter or separate the water with another column before letting the sample enter the molecular sieve, but honestly my findings were rather sparse this far. Do you know of any new findings or is that post still up to date?

    Thanks in advance and the great post in general!

    Regards,
    Sanni.

  21. Alan Sensue says:

    To our blog post readers, please note that before any of us blog post authors are permitted to answer any questions, we need to obtain your contact information (your full name, company name and mailing address). I emailed Sanni to obtain this information, which is why there was a delay in my reply/response. Please note that your full name, company name and mailing address will not be posted with the comment/question. Thank you.

    Hello Sanni: Unfortunately, I am not aware of any advancements with molecular sieve materials and/or columns since I wrote this post several years ago. However, there have been some advances in instrumentation (smaller units, faster analysis, more sensitivity), but certain set-up principles still apply using GCs containing valves and multiple columns (for extended compound range analysis and/or back-flushing). You may want to review this other post on the topic I wrote. https://blog.restek.com/?p=14370

    I hope this helps answer your questions. Regards. Alan

  22. Mostafa Aly says:

    Hello guys,

    Lots of valuable information on this blog! Thanks to everyone who has contributed so far!

    I’m currently seeing a small oxygen and nitrogen peak, followed by a smaller peak from my TCD detector (Molsieve column). The oxygen peak is larger than the nitrogen peak and that makes me think it’s not a leak. In addition, we have not changed the purities of the gases we use i.e. not sure if its impurities. The third peak after the nitrogen is not methane as I can clearly see both methane and carbon monoxide eluting at later times.

    Does anyone have any idea why I’m seeing these peaks? And what is this third peak after nitrogen and before methane?
    Thanks in advance!
    Regards,
    Mostafa

  23. Alan Sensue says:

    Hello: As you know, in order for a column to chromatograph compound peaks, the compounds need to be introduced to the head of the column. Once they reach the head of the column, they can either condense and re-vaporize when the temperature is hot enough, or they can enter into the column without condensing and travel the entire length of the column. Because O2 and N2 are the primary compounds in question, these are not going to condense at the head of the column. In summary, you should not see these particular compounds/gases as “ghost peaks”. To look at this another way, a steady leak of air into the carrier gas or detector may cause an elevated TCD (baseline) signal, but should not cause distinct peaks.

    I think before anyone can guess as to which compound(s) may be in your chromatogram, more will need to be known about your samples/standards, instrument and the analytical/method parameters. Are these peaks there with every injection at the same retention time? If so, they are likely from whatever you are injecting. If they show up at random times, they are likely impurities which eventually elute from the column, especially if no injection is made. In addition, please keep in mind that some compounds, like SF6, will not fit into the pores and will quickly exit a 5A column, but this compound would elute before nitrogen. If the pores of the material are filled with moisture and CO2, carbon monoxide could elute between nitrogen and methane, but I do not know which other gases may elute in this area too.

    I hope this helps. If you continue to experience difficulties with this analysis, feel free to contact support via https://www.restek.com/Contact-Us/Technical-Service

    Alan

  24. CARL JASTRZEMBSKI says:

    I have a question I hope someone can answer. I am using an Agilent 7890B with a helium carrier. It has a silica gel pre column 60/80, mole sieve 13x 80/100 column. my question is with the order of elution. These peaks were named before I came here but I dont think they are right. it shows in this order H2, Ar, N2, Kr, CH4, CO.
    I dont think the column can separate the Kr from N2 but I could be wrong. Also the sample has Neon and CO2 in it but im not sure if it would show this because of the Pre column and back flush

  25. Alan Sensue says:

    Hi Carl:

    Thank you for reading my post. Unfortunately, I do not have any personal experience using Silica Gel as the pre-column (stripper/back-flush column), so I do not know how it would affect the elution order of compounds/gases. What I do know is that the Silica Gel column is good at retaining moisture and CO2, which is why it is a common pre-column when using a molecular sieve as the primary column (like a 5A or 13X). In addition, CO2 would be retained on the 13X, so you would likely not detect a peak with your current set-up unless CO2 levels were in the double-digit % levels or higher, and values would likely be non-reproducible.

    Based upon information I was able to find (and by no means does it completely answer your question), compound elution order for a Silica Gel and MSieve 13X column are listed below.

    Silica Gel: hydrogen, air, carbon monoxide, methane, ethane, carbon dioxide

    Molecular Sieve 13X: helium/neon, hydrogen, argon, oxygen, nitrogen/krypton*, methane, carbon monoxide. *Please note I had found chromatograms where nitrogen and krypton co-eluted, where nitrogen eluted before krypton, and even one where nitrogen eluted after krypton. However, none were Restek chromatograms so I cannot verify their accuracy. But, nitrogen eluting before kryton seemed to be the most common elution order.

    I hope this helps.

    Alan

  26. GYAN BATRA says:

    This is regarding TOGA application.
    Most of Data published have used more then one packed column with Gas Sampling Valve for column switching.
    I have done this application using S.S. Column packed with either Carbosphere or Carbosieve S II connected to TCD, Methanizer and FID connected in series. Dual Channel Chromatography Data Station is used. Oven Temp. 70 Degree C for 1 Min. Temp.Programming @15 Degree C/ Min. Final Temp. 220 Degree C. Sequence of elution is Hydrogen on TCD and CO, CH4, CO2, C2H2, C2H4 & C2H6 on FID
    This makes analysis very simple and result obtained are quite good.
    Pl let me know if I can use ShinCarbon ST column material for this application.

  27. Alan Sensue says:

    Hello: Thank you for your comment/question.

    Unfortunately, we have no comparison data between the ShinCarbon and any other carbon molecular sieve materials. For GCs with single columns, substitution of the ShinCarbon for a different carbon molecular sieve column should be fairly simple, but if part of a process GC, or when multiple columns are involved, a simple substitution may not be possible. In this case, the GC valve timing, temperatures and flow rates may need to be adjusted to provide similar chromatography.

    Regards, Alan

  28. Molly Augustine says:

    Hello,

    When you stated “helium (at high temperatures) can remove the oxidation layer on the metal’s surface, which can lead to activity (especially for CO)” … what temperature range would you consider high? And below what temperature would you estimate there would be no effect.

    Thanks, Molly

  29. Alan Sensue says:

    Hi Molly. You asked a good question, which I do not know the answer to, and as you inferred, I can only guess. I my mind, I think of 250°C and higher as the “high” temperature above which the reaction with the particle’s surface that with certain gases, especially hydrogen, will strip the oxidation layer off the metal surface leaving a reactive particle. Honestly, I think of this temperature (250°C) because between 250°C and 350°C is the temperature range used to dry the column (using dry, clean nitrogen as the carrier gas) and/or to regenerate/reactivate the column (using dry, clean air as the carrier gas), and it’s during this temperature range (250°C to 350°C) that reactivity issues occur (I’ve never seen low temperatures, like 40°C, cause any reactivity issues with these columns).

    I hope you find this helpful (even if there is no data to back it up). Thank you for reading my post. Alan

  30. Mostafa Aly says:

    Thank you Alan for all the replies!

    I just installed a new MolSieve 5A on an RGA setup and my CO is eluting very late compared to my previous one!
    It is currently eluting at around 9 mins whereas it used to elute at around 6 min. The peak is also tailing a lot thus not so symmetric.

    Do you have any idea why this could be happening?

    PS: CH4 is eluting at 6 min instead of 5.8 min previously. Seems like the gap between CH4 and CO just got much bigger.

  31. Alan Sensue says:

    Hello Mostafa. There are two primary reasons that I can think of which may cause the issue you have observed concerning retention times.

    1) You have not set the carrier gas flow as described in this link. https://blog.restek.com/?p=13257

    2) When the pores in the particles are free of moisture & CO2 and surface contamination is minimal, the separation between all compounds, especially between methane and carbon monoxide, will increase.

    I hope this helps. Alan

  32. Jay Lee says:

    Hello.

    I have very basic question.

    What is difference between porous polyer and molsieve(molecular Sieve)?

    I found that Molsieve is zeolite or active carbon, active glass etc.
    Then what is porous polymer? (Like Hayesep, porapak)

    Inorganic compound and Polymer is that a only difference?

    Thanks , Jay Lee

  33. Alan Sensue says:

    Hello: I think the best way for me to show you the differences is for you to review the links below. I hope they explain each type of packing to your satisfaction. If you need additional information, you may want to contact a manufacturer of the specific packing you are interested in learning more about. As a separations specialist, I am more concerned which applications they are recommended for rather than any physical or chemical properties. But generally speaking, your assessment of the differences is accurate.

    I hope this helps. Alan

    Carbon Molecular Sieves https://www.sciencedirect.com/topics/chemistry/carbon-molecular-sieve

    Zeolite Molecular Sieves https://en.wikipedia.org/wiki/Molecular_sieve

    Porous polymers https://www.sciencedirect.com/topics/chemistry/porous-polymer

  34. Yuan says:

    Hi Alan,

    For PLOT molecular 5A column, we saw adsorption of oxygen under low levels (eg, 5ppm). Argon/oxygen separation is needed in our application. But it seems our PLOT 5A column absorb oxygen and we only get an Argon peak. Our mixture is about 5 ppm or 10 ppm. We tried different brand of 5A PLOT columns, but both got no oxygen peak. And we tried to condition the column with air or nitrogen, but it didn’t work. We are still struggling on this. Do you know some reliable manufactures who can provide PLOT 5A column without oxygen absorption issue? Thanks.

  35. Alan Sensue says:

    Hello Yuan: Do you know what levels of oxygen your instrument is capable of detecting?

    I have a few questions for you:

    1. What detector are you using? Are you absolutely certain that it can detect these low levels of oxygen?

    2. Because separation of oxygen and argon is so difficult, I would suggest using a different gas standard for troubleshooting, one that ideally only contains oxygen in a balance gas of helium.

    3. Are there any GC columns upstream of the molecular sieve column? Porous polymers (like HayeSep, Porapak, Chromosorb polymers) are known to absorb low levels of oxygen.

    Please provide additional details about your instrument, detector, and if any other columns are also used for this analysis and hopefully we will be able to determine the reason you are not able to detect oxygen at these levels.

    Alan

Leave a Reply


Restek Domestic Customer Service

Subject

Message