What are these O-rings for that I received with my Super-Clean baseplate filter/trap?

Some of you have noticed that included with several of our baseplate filters/traps are a package of two small O-rings. Technical Service has been informed that these are the O-rings which fit onto the check valves on your baseplate (see red arrows below).  Does this mean that you need to replace these O-rings?  Not necessarily.  Carefully inspect the O-rings on the check valves, and if you do not see any signs of wear (cracks, deformation, etc…), you probably don’t need to replace them.  However, if they do exhibit signs of wear, go ahead and replace them.

.

 Baseplate 4

.

If you also need the larger O-rings that are located on the baseplate, at the base of the check valves (as identified by the green arrows), you can obtain those by purchasing Restek catalog number 22023.

22023